{"id":"CVE-2014-5406","aliases":["EUVD-2014-5294","GHSA-vxf4-qh89-w2r6"],"status":"published","title":"Hospira LifeCare PCA Infusion System","description":"The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.  NOTE: this issue might overlap CVE-2015-3459.","severity":{"version":"2.0","score":7.6,"vector":"AV:N/AC:H/Au:N/C:C/I:C/A:C","assessor":"cna","from":"cve"},"epss":{"day":"2026-11-07","score":0.01218,"percentile":0.68807,"model":"v2026.06.15"},"facts":{"in_kev":false,"known_exploited":false,"public_exploit":false,"fix_available":false},"cwes":["CWE-345"],"published":"2015-07-06T19:10:00Z","modified":"2026-06-17T00:11:30Z","products":[{"vendor":"hospira","product":"lifecare pca infusion system"},{"vendor":"hospira","product":"lifecare_pcainfusion_firmware"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01"},{"url":"http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm","tags":["x_refsource_MISC","x_transferred","Third Party Advisory","US Government Resource"]},{"url":"https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/","tags":["x_refsource_MISC","x_transferred"]},{"url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json"},{"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01","tags":["x_refsource_MISC","x_transferred","Third Party Advisory","US Government Resource"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2014-5406"},{"url":"https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities"},{"url":"https://github.com/advisories/GHSA-vxf4-qh89-w2r6"}],"sources":{"cve":[{"id":"CVE-2014-5406","title":"Hospira LifeCare PCA Infusion System","description":"The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.  NOTE: this issue might overlap CVE-2015-3459.","status":"published","published":"2015-07-06T19:10:00Z","modified":"2025-11-03T18:34:36.324Z","cvss":[{"version":"2.0","score":7.6,"vector":"AV:N/AC:H/Au:N/C:C/I:C/A:C","assessor":"cna"}],"cwes":["CWE-345"],"updated_at":"2026-11-02T14:33:42.219637Z"}],"euvd":[{"id":"EUVD-2014-5294","description":"The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.  NOTE: this issue might overlap CVE-2015-3459.","status":"published","published":"2015-07-06T19:10:00Z","modified":"2025-11-03T18:34:36Z","cvss":[{"version":"2.0","score":7.6,"vector":"AV:N/AC:H/Au:N/C:C/I:C/A:C","level":"high"}],"updated_at":"2026-11-02T19:49:34.635725Z"}],"ghsa":[{"id":"GHSA-vxf4-qh89-w2r6","title":"The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated...","description":"The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.  NOTE: this issue might overlap CVE-2015-3459.","status":"published","published":"2022-05-17T04:11:59Z","modified":"2025-11-03T21:30:29Z","cwes":["CWE-345"],"updated_at":"2026-11-02T15:04:18.176646Z"}],"nvd":[{"id":"CVE-2014-5406","description":"The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port.  NOTE: this issue might overlap CVE-2015-3459.","status":"published","published":"2015-07-06T19:59:00.097Z","modified":"2026-06-17T00:11:30.157Z","cvss":[{"version":"2.0","score":7.6,"vector":"AV:N/AC:H/Au:N/C:C/I:C/A:C","level":"high","assessor":"secondary"},{"version":"2.0","score":9.3,"vector":"AV:N/AC:M/Au:N/C:C/I:C/A:C","level":"high","assessor":"primary"}],"cwes":["CWE-345"],"updated_at":"2026-11-02T14:02:58.385843Z"}]}}