{"id":"CVE-2018-10601","aliases":["EUVD-2018-2673","GHSA-9f6x-r29g-cv4j"],"status":"published","title":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M,...","description":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an \"echo\" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.","severity":{"version":"3.1","score":8.2,"vector":"CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H","level":"high","assessor":"primary","from":"nvd"},"epss":{"day":"2026-11-08","score":0.00372,"percentile":0.31343,"model":"v2026.06.15"},"facts":{"in_kev":false,"known_exploited":false,"public_exploit":false,"fix_available":false},"cwes":["CWE-121","CWE-787"],"published":"2018-06-05T20:00:00Z","modified":"2026-06-17T01:34:15Z","products":[{"vendor":"ics-cert","product":"intellivue patient monitors, avalon fetal/maternal monitors"},{"vendor":"philips","product":"avalon_fetal\\/maternal_monitors_fm20_firmware"},{"vendor":"philips","product":"avalon_fetal\\/maternal_monitors_fm30_firmware"},{"vendor":"philips","product":"avalon_fetal\\/maternal_monitors_fm40_firmware"},{"vendor":"philips","product":"avalon_fetal\\/maternal_monitors_fm50_firmware"},{"vendor":"philips","product":"intellivue_mp2_firmware"},{"vendor":"philips","product":"intellivue_mp30_firmware"},{"vendor":"philips","product":"intellivue_mp50_firmware"},{"vendor":"philips","product":"intellivue_mp70_firmware"},{"vendor":"philips","product":"intellivue_mx100_firmware"},{"vendor":"philips","product":"intellivue_mx400_firmware"},{"vendor":"philips","product":"intellivue_mx450_firmware"},{"vendor":"philips","product":"intellivue_mx500_firmware"},{"vendor":"philips","product":"intellivue_mx550_firmware"},{"vendor":"philips","product":"intellivue_mx700_firmware"},{"vendor":"philips","product":"intellivue_mx800_firmware"},{"vendor":"philips","product":"intellivue_np90_firmware"},{"vendor":"philips","product":"intellivue_x2_firmware"},{"vendor":"philips","product":"intellivue_x3_firmware"}],"references":[{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01","tags":["x_refsource_MISC","x_transferred","Third Party Advisory","US Government Resource"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2018-10601"},{"url":"https://github.com/advisories/GHSA-9f6x-r29g-cv4j"}],"sources":{"cve":[{"id":"CVE-2018-10601","description":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an \"echo\" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.","status":"published","published":"2018-06-05T20:00:00Z","modified":"2024-09-17T02:36:04.782Z","cwes":["CWE-121"],"updated_at":"2026-11-02T14:35:01.28843Z"}],"euvd":[{"id":"EUVD-2018-2673","description":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an \"echo\" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.","status":"published","published":"2018-06-05T20:00:00Z","modified":"2024-09-17T02:36:04Z","updated_at":"2026-11-03T00:28:39.162217Z"}],"ghsa":[{"id":"GHSA-9f6x-r29g-cv4j","title":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M,...","description":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an \"echo\" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.","status":"published","published":"2022-05-13T01:10:22Z","modified":"2023-02-01T05:03:36Z","cvss":[{"version":"3.1","score":8.2,"vector":"CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H","level":"high"}],"cwes":["CWE-787"],"updated_at":"2026-11-02T14:50:05.01868Z"}],"nvd":[{"id":"CVE-2018-10601","description":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an \"echo\" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.","status":"published","published":"2018-06-05T20:29:00.903Z","modified":"2026-06-17T01:34:15.347Z","cvss":[{"version":"3.1","score":8.2,"vector":"CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H","level":"high","assessor":"primary"},{"version":"2.0","score":5.4,"vector":"AV:A/AC:M/Au:N/C:P/I:P/A:P","level":"medium","assessor":"primary"}],"cwes":["CWE-121","CWE-787"],"updated_at":"2026-11-02T14:04:28.483469Z"}]}}