{"id":"CVE-2019-25722","aliases":["EUVD-2019-20158","GHSA-fh3f-j952-25x6"],"status":"published","title":"Dräger SC Monitoring Devices Hard-coded Credentials and DoS","description":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.","severity":{"version":"4.0","score":7.2,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","level":"high","assessor":"cna","from":"cve"},"epss":{"day":"2026-11-10","score":0.00193,"percentile":0.09037,"model":"v2026.06.15"},"facts":{"in_kev":false,"known_exploited":false,"public_exploit":false,"fix_available":false},"cwes":["CWE-798"],"published":"2026-06-02T18:44:11Z","modified":"2026-07-22T19:10:00Z","products":[{"vendor":"dräger","product":"sc 6002xl"},{"vendor":"dräger","product":"sc 7000"},{"vendor":"dräger","product":"sc6802xl"},{"vendor":"dräger","product":"sc8000"},{"vendor":"dräger","product":"sc90000 xl"}],"references":[{"url":"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-hard-coded-credentials-and-dos","tags":["third-party-advisory"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2019-25722"},{"url":"https://static.draeger.com/security"},{"url":"https://github.com/advisories/GHSA-fh3f-j952-25x6"}],"sources":{"cve":[{"id":"CVE-2019-25722","title":"Dräger SC Monitoring Devices Hard-coded Credentials and DoS","description":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.","status":"published","published":"2026-06-02T18:44:11.815Z","modified":"2026-06-03T16:19:46.284Z","cvss":[{"version":"4.0","score":7.2,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","level":"high","assessor":"cna"},{"version":"3.1","score":7.6,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","level":"high","assessor":"cna"}],"cwes":["CWE-798"],"updated_at":"2026-11-02T14:20:19.399352Z"}],"euvd":[{"id":"EUVD-2019-20158","description":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.","status":"published","published":"2026-06-02T18:44:11Z","modified":"2026-06-03T16:19:46Z","cvss":[{"version":"4.0","score":7.2,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N","level":"high"}],"updated_at":"2026-11-02T18:20:53.780241Z"}],"ghsa":[{"id":"GHSA-fh3f-j952-25x6","title":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard...","description":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.","status":"published","published":"2026-06-02T21:30:41Z","modified":"2026-06-03T18:34:14Z","cvss":[{"version":"4.0","score":7.2,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","level":"high"},{"version":"3.1","score":7.6,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","level":"high"}],"cwes":["CWE-798"],"updated_at":"2026-11-02T15:08:31.387226Z"}],"nvd":[{"id":"CVE-2019-25722","description":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.","status":"published","published":"2026-06-02T20:16:23.69Z","modified":"2026-07-22T19:10:00.12Z","cvss":[{"version":"4.0","score":7.2,"vector":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X","level":"high","assessor":"secondary"},{"version":"3.1","score":7.6,"vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H","level":"high","assessor":"secondary"}],"cwes":["CWE-798"],"updated_at":"2026-11-02T14:05:50.275357Z"}]}}