{"id":"CVE-2020-12036","aliases":["EUVD-2020-4352","GHSA-cx3p-ghw7-pr4c"],"status":"published","title":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not...","description":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.","severity":{"version":"3.1","score":7.5,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N","level":"high","assessor":"primary","from":"nvd"},"epss":{"day":"2026-11-12","score":0.00496,"percentile":0.43062,"model":"v2026.06.15"},"facts":{"in_kev":false,"known_exploited":false,"public_exploit":false,"fix_available":false},"cwes":["CWE-319"],"published":"2020-06-29T13:49:53Z","modified":"2026-06-17T02:51:18Z","products":[{"vendor":"baxter","product":"prismaflex_firmware"},{"vendor":"baxter","product":"prismax_firmware"}],"references":[{"url":"https://www.us-cert.gov/ics/advisories/icsma-20-170-01","tags":["x_refsource_MISC","x_transferred","Not Applicable"]},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2020-12036"},{"url":"https://github.com/advisories/GHSA-cx3p-ghw7-pr4c"}],"sources":{"cve":[{"id":"CVE-2020-12036","description":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.","status":"published","published":"2020-06-29T13:49:53Z","modified":"2024-08-04T11:48:57.981Z","cwes":["CWE-319"],"updated_at":"2026-11-02T14:35:52.925152Z"}],"euvd":[{"id":"EUVD-2020-4352","description":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.","status":"published","published":"2020-06-29T13:49:53Z","modified":"2024-08-04T11:48:57Z","updated_at":"2026-11-03T03:07:41.399649Z"}],"ghsa":[{"id":"GHSA-cx3p-ghw7-pr4c","title":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not...","description":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.","status":"published","published":"2022-05-24T17:21:59Z","modified":"2023-01-29T05:03:32Z","updated_at":"2026-11-02T14:45:20.476531Z"}],"nvd":[{"id":"CVE-2020-12036","description":"Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.","status":"published","published":"2020-06-29T14:15:11.46Z","modified":"2026-06-17T02:51:18.673Z","cvss":[{"version":"3.1","score":7.5,"vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N","level":"high","assessor":"primary"},{"version":"2.0","score":5,"vector":"AV:N/AC:L/Au:N/C:P/I:N/A:N","level":"medium","assessor":"primary"}],"cwes":["CWE-319"],"updated_at":"2026-11-02T14:06:08.145904Z"}]}}