{"id":"CVE-2024-48971","aliases":["EUVD-2024-43156","GHSA-78wf-p97g-f4jr"],"status":"published","title":"Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator","description":"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.","severity":{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical","assessor":"cna","from":"cve"},"epss":{"day":"2026-11-09","score":0.00216,"percentile":0.12186,"model":"v2026.06.15"},"facts":{"in_kev":false,"known_exploited":false,"public_exploit":false,"fix_available":false},"cwes":["CWE-798"],"published":"2024-11-14T21:13:36Z","modified":"2026-06-17T07:59:08Z","products":[{"vendor":"baxter","product":"life2000 ventilation system"},{"vendor":"baxter","product":"life2000_ventilator_firmware"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2024-48971"},{"url":"https://github.com/advisories/GHSA-78wf-p97g-f4jr"}],"sources":{"cve":[{"id":"CVE-2024-48971","title":"Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator","description":"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.","status":"published","published":"2024-11-14T21:13:36.036Z","modified":"2024-11-15T21:06:24.325Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical","assessor":"cna"}],"cwes":["CWE-798"],"updated_at":"2026-11-02T14:34:11.97482Z"}],"euvd":[{"id":"EUVD-2024-43156","description":"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.","status":"published","published":"2024-11-14T21:13:36Z","modified":"2024-11-15T21:06:24Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical"}],"updated_at":"2026-11-03T00:04:57.661321Z"}],"ghsa":[{"id":"GHSA-78wf-p97g-f4jr","title":"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in...","description":"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.","status":"published","published":"2024-11-15T00:31:51Z","modified":"2024-11-15T00:31:58Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical"}],"cwes":["CWE-798"],"updated_at":"2026-11-02T14:58:50.830346Z"}],"nvd":[{"id":"CVE-2024-48971","description":"The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.","status":"published","published":"2024-11-14T22:15:18.327Z","modified":"2026-06-17T07:59:08.533Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical","assessor":"secondary"}],"cwes":["CWE-798"],"updated_at":"2026-11-02T14:10:37.281254Z"}]}}