{"id":"CVE-2024-48973","aliases":["EUVD-2024-43157","GHSA-xx59-963p-2v9w"],"status":"published","title":"Debug port on Life2000 Ventilator serial interface is enabled by default","description":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.","severity":{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical","assessor":"cna","from":"cve"},"epss":{"day":"2026-11-09","score":0.0022,"percentile":0.12639,"model":"v2026.06.15"},"facts":{"in_kev":false,"known_exploited":false,"public_exploit":false,"fix_available":false},"cwes":["CWE-1263"],"published":"2024-11-14T21:24:14Z","modified":"2026-06-17T07:59:08Z","products":[{"vendor":"baxter","product":"life2000 ventilation system"},{"vendor":"baxter","product":"life2000_ventilator_firmware"}],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"},{"url":"https://nvd.nist.gov/vuln/detail/CVE-2024-48973"},{"url":"https://github.com/advisories/GHSA-xx59-963p-2v9w"}],"sources":{"cve":[{"id":"CVE-2024-48973","title":"Debug port on Life2000 Ventilator serial interface is enabled by default","description":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.","status":"published","published":"2024-11-14T21:24:14.099Z","modified":"2024-11-18T15:33:44.924Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical","assessor":"cna"}],"cwes":["CWE-1263"],"updated_at":"2026-11-02T14:34:11.97482Z"}],"euvd":[{"id":"EUVD-2024-43157","description":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.","status":"published","published":"2024-11-14T21:24:14Z","modified":"2024-11-18T15:33:44Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical"}],"updated_at":"2026-11-03T00:04:57.054288Z"}],"ghsa":[{"id":"GHSA-xx59-963p-2v9w","title":"The debug port on the ventilator's serial interface is enabled by default. This could allow an...","description":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.","status":"published","published":"2024-11-15T00:31:51Z","modified":"2024-11-15T00:31:58Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical"}],"cwes":["CWE-1263"],"updated_at":"2026-11-02T14:58:50.830346Z"}],"nvd":[{"id":"CVE-2024-48973","description":"The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.","status":"published","published":"2024-11-14T22:15:18.51Z","modified":"2026-06-17T07:59:08.633Z","cvss":[{"version":"3.1","score":9.3,"vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","level":"critical","assessor":"secondary"}],"cwes":["CWE-1263"],"updated_at":"2026-11-02T14:10:37.281254Z"}]}}