Every CVE Search/ Get API key

CVE-2014-5431

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module ...

6.8 Medium

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Attack profile

How easy it is to exploit

Attack vector
Physical
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.38%Down 0.5 points since Feb 4, 2022
90%50%10%1%

What each source reports

NVD Score used 6.8

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Record
CVE-2014-5431
CVSS 3.0
6.8 MediumCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
4.6 MediumAV:L/AC:L/Au:N/C:P/I:P/A:P
Weakness
CWE-259, CWE-798
Updated
2026-11-02
CVE List No score

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Record
CVE-2014-5431
Weakness
CWE-259
Updated
2026-11-02
EUVDEUVD-2014-5319 No score

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Record
EUVD-2014-5319
Updated
2026-11-03
GitHub AdvisoriesGHSA-2wv9-4g7m-79pc 6.8

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module ...

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Record
GHSA-2wv9-4g7m-79pc
CVSS 3.0
6.8 MediumCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-798
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2014-5431 is a medium vulnerability (CVSS 6.8) in Baxter Sigma Spectrum Infusion System, published March 26, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.38%, the 31st percentile. No source names a fixed version.

References

MoveEnter Open