Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System....

7.5 High

CVE-2015-1012 is a high vulnerability (CVSS 7.5) in Hospira Lifecare Pca Infusion System, published March 25, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.77%, the 55th percentile. No source names a fixed version.

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

0.77%Down 0.1 points since Feb 4, 2022
90%50%10%1%

What each source reports

NVD Score used 7.5

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Record
CVE-2015-1012
CVSS 3.0
7.5 HighCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0
5.0 MediumAV:N/AC:L/Au:N/C:P/I:N/A:N
Weakness
CWE-312, CWE-200
Updated
2026-11-02
CVE List No score

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Record
CVE-2015-1012
Weakness
CWE-312
Updated
2026-11-02
EUVDEUVD-2015-1156 No score

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Record
EUVD-2015-1156
Updated
2026-11-03
GitHub AdvisoriesGHSA-82mj-h4mj-5q4v 7.5

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System....

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Record
GHSA-82mj-h4mj-5q4v
CVSS 3.0
7.5 HighCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Weakness
CWE-200
Updated
2026-11-02

Affected products

History

History
WhenChangedSources
2026-11-03aliasesEUVD
2026-11-02createdCVE List, GitHub Advisories, NVD

References

MoveEnter Open