Every CVE Search/ Get API key

CVE-2015-3953

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior,...

9.8 Critical

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

1.97%Up 1.1 points since Feb 4, 2022
90%50%10%1%

What each source reports

NVD Score used 9.8

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Record
CVE-2015-3953
CVSS 3.0
9.8 CriticalCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
10.0 HighAV:N/AC:L/Au:N/C:C/I:C/A:C
Weakness
CWE-259, CWE-798
Updated
2026-11-02
CVE List No score

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Record
CVE-2015-3953
Weakness
CWE-259
Updated
2026-11-02
EUVDEUVD-2015-3984 No score

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Record
EUVD-2015-3984
Updated
2026-11-03
GitHub AdvisoriesGHSA-vw52-2pgm-8cj2 9.8

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior,...

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Record
GHSA-vw52-2pgm-8cj2
CVSS 3.0
9.8 CriticalCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-798
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2015-3953 is a critical vulnerability (CVSS 9.8) in Hospira Plum A+ Infusion System, published March 25, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.97%, the 80th percentile. No source names a fixed version.

References

MoveEnter Open