Every CVE Search/ Get API key

CVE-2016-8355

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0...

9.9 Critical

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
Low
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

1.49%Up 0.6 points since Feb 4, 2022
90%50%10%1%

What each source reports

NVD Score used 9.9

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.

Record
CVE-2016-8355
CVSS 3.0
9.9 CriticalCVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
CVSS 2.0
9.0 HighAV:N/AC:L/Au:S/C:C/I:C/A:C
Weakness
CWE-306
Updated
2026-11-02
CVE List No score

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.

Record
CVE-2016-8355
Updated
2026-11-02
EUVDEUVD-2016-9203 No score

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.

Record
EUVD-2016-9203
Updated
2026-11-03
GitHub AdvisoriesGHSA-2h29-mg3v-85xj 9.9

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0...

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.

Record
GHSA-2h29-mg3v-85xj
CVSS 3.0
9.9 CriticalCVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-306
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2016-8355 is a critical vulnerability (CVSS 9.9) in Smiths-medical Cadd-solis Medication Safety Software, published February 13, 2017. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.49%, the 74th percentile. No source names a fixed version.

References

MoveEnter Open