Every CVE Search/ Get API key

CVE-2016-8358

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0...

8.5 High

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
High
Privileges needed
Low
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.50%Down 0.4 points since Feb 4, 2022
90%50%10%1%

What each source reports

NVD Score used 8.5

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.

Record
CVE-2016-8358
CVSS 3.0
8.5 HighCVSS:3.0/AV:N/AC:H/PR:L/UI:N/S:C/C:H/I:H/A:H
CVSS 2.0
6.0 MediumAV:N/AC:M/Au:S/C:P/I:P/A:P
Weakness
CWE-346
Updated
2026-11-02
CVE List No score

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.

Record
CVE-2016-8358
Updated
2026-11-02
EUVDEUVD-2016-9206 No score

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.

Record
EUVD-2016-9206
Updated
2026-11-03
GitHub AdvisoriesGHSA-6c6r-7283-r9ph 8.5

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0...

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.

Record
GHSA-6c6r-7283-r9ph
CVSS 3.0
8.5 HighCVSS:3.0/AV:N/AC:H/PR:L/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-346
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2016-8358 is a high vulnerability (CVSS 8.5) in Smiths-medical Cadd-solis Medication Safety Software, published February 13, 2017. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.50%, the 43rd percentile. No source names a fixed version.