Every CVE Search/ Get API key

CVE-2017-11578

It was discovered as a part of the research on IoT devices in the most recent firmware for...

5.9 Medium

It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
High
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

2.05%Up 1.2 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 5.9

It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

Record
CVE-2017-11578
CVSS 3.0
5.9 MediumCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0
4.3 MediumAV:N/AC:M/Au:N/C:P/I:N/A:N
Weakness
CWE-200
Updated
2026-11-02
CVE List No score

It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

Record
CVE-2017-11578
Updated
2026-11-02
EUVDEUVD-2017-3193 No score

It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

Record
EUVD-2017-3193
Updated
2026-11-03
GitHub AdvisoriesGHSA-hjx2-43ph-gqjc 5.9

It was discovered as a part of the research on IoT devices in the most recent firmware for...

It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

Record
GHSA-hjx2-43ph-gqjc
CVSS 3.0
5.9 MediumCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
Weakness
CWE-200
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2017-11578 is a medium vulnerability (CVSS 5.9) in Blipcare Wi-fi Blood Pressure Monitor Firmware, published July 2, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 2.05%, the 81st percentile. No source names a fixed version.

References

MoveEnter Open