CVE-2017-12716
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit...
CVE-2017-12716 is a medium vulnerability (CVSS 6.5) in Abbott Laboratories Accent And Anthem, published April 25, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.24%, the 15th percentile. No source names a fixed version.
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Attack profile
How easy it is to exploit
- Attack vector
- Adjacent network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- None
- Availability
- None
Exploit likelihood
What each source reports
NVD Score used
6.5
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
- Record
CVE-2017-12716- CVSS 3.0
- 6.5 Medium
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N - CVSS 2.0
- 3.3 Low
AV:A/AC:L/Au:N/C:P/I:N/A:N - Weakness
- CWE-311, CWE-319
- Updated
- 2026-11-02
CVE List No score
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
- Record
CVE-2017-12716- Weakness
- CWE-311
- Updated
- 2026-11-02
EUVDEUVD-2017-4255 No score
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
- Record
EUVD-2017-4255- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-x247-46rv-347w
6.5
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit...
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
- Record
GHSA-x247-46rv-347w- CVSS 3.0
- 6.5 Medium
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N - Weakness
- CWE-319
- Updated
- 2026-11-02
Affected products
History
| When | Changed | Sources |
|---|---|---|
| 2026-11-03 | aliases | EUVD |
| 2026-11-02 | created | CVE List, GitHub Advisories, NVD |