CVE-2017-12718
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe...
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- High
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
NVD Score used
8.1
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
- Record
CVE-2017-12718- CVSS 3.0
- 8.1 High
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H - CVSS 2.0
- 6.8 Medium
AV:N/AC:M/Au:N/C:P/I:P/A:P - Weakness
- CWE-120, CWE-119
- Updated
- 2026-11-02
CVE List No score
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
- Record
CVE-2017-12718- Weakness
- CWE-120
- Updated
- 2026-11-02
EUVDEUVD-2017-4257 No score
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
- Record
EUVD-2017-4257- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-84pw-5frw-ww2p
8.1
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe...
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
- Record
GHSA-84pw-5frw-ww2p- CVSS 3.0
- 8.1 High
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-119
- Updated
- 2026-11-02
Affected products
Public exploits
Links only. Every CVE stores no exploit code.
| Source | Link | Date |
|---|---|---|
| Exploit-DB | Smiths Medical Medfusion 4000 - 'DHCP' Denial of Service | 2018-01-18 |
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, Exploit-DB, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2017-12718 is a high vulnerability (CVSS 8.1) in Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump, published February 15, 2018. Public exploit code exists. EPSS puts the chance of exploitation in the next 30 days at 12.8%, the 96th percentile. No source names a fixed version.