CVE-2017-12718

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe...

8.1 HighPublic exploit

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
High
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

12.8%Up 3.3 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 8.1

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Record
CVE-2017-12718
CVSS 3.0
8.1 HighCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
6.8 MediumAV:N/AC:M/Au:N/C:P/I:P/A:P
Weakness
CWE-120, CWE-119
Updated
2026-11-02
CVE List No score

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Record
CVE-2017-12718
Weakness
CWE-120
Updated
2026-11-02
EUVDEUVD-2017-4257 No score

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Record
EUVD-2017-4257
Updated
2026-11-03
GitHub AdvisoriesGHSA-84pw-5frw-ww2p 8.1

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe...

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Record
GHSA-84pw-5frw-ww2p
CVSS 3.0
8.1 HighCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-119
Updated
2026-11-02

Affected products

Public exploits

Links only. Every CVE stores no exploit code.

Public exploits
SourceLinkDate
Exploit-DBSmiths Medical Medfusion 4000 - 'DHCP' Denial of Service2018-01-18

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, Exploit-DB, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2017-12718 is a high vulnerability (CVSS 8.1) in Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump, published February 15, 2018. Public exploit code exists. EPSS puts the chance of exploitation in the next 30 days at 12.8%, the 96th percentile. No source names a fixed version.