CVE-2017-12722
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe...
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- None
- Integrity
- None
- Availability
- Low
Exploit likelihood
What each source reports
NVD Score used
5.3
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
- Record
CVE-2017-12722- CVSS 3.0
- 5.3 Medium
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L - CVSS 2.0
- 5.0 Medium
AV:N/AC:L/Au:N/C:N/I:N/A:P - Weakness
- CWE-125
- Updated
- 2026-11-02
CVE List No score
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
- Record
CVE-2017-12722- Updated
- 2026-11-02
EUVDEUVD-2017-4261 No score
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
- Record
EUVD-2017-4261- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-92p5-8hff-v96g
5.3
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe...
An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
- Record
GHSA-92p5-8hff-v96g- CVSS 3.0
- 5.3 Medium
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L - Weakness
- CWE-125
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2017-12722 is a medium vulnerability (CVSS 5.3) in Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump, published February 15, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 2.42%, the 84th percentile. No source names a fixed version.