Every CVE Search/ Get API key

CVE-2017-12725

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless...

5.6 Medium

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
High
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
Low
Integrity
Low
Availability
Low

Exploit likelihood

0.83%Up 0.2 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 5.6

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Record
CVE-2017-12725
CVSS 3.0
5.6 MediumCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
CVSS 2.0
6.8 MediumAV:N/AC:M/Au:N/C:P/I:P/A:P
Weakness
CWE-798
Updated
2026-11-02
CVE List No score

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Record
CVE-2017-12725
Updated
2026-11-02
EUVDEUVD-2017-4264 No score

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Record
EUVD-2017-4264
Updated
2026-11-03
GitHub AdvisoriesGHSA-chgj-p4c2-8wqj 5.6

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless...

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Record
GHSA-chgj-p4c2-8wqj
CVSS 3.0
5.6 MediumCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
Weakness
CWE-798
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2017-12725 is a medium vulnerability (CVSS 5.6) in Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump, published February 15, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.83%, the 57th percentile. No source names a fixed version.

References

MoveEnter Open