A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless...
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- High
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- Low
- Integrity
- Low
- Availability
- Low
Exploit likelihood
What each source reports
NVD Score used
5.6
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
- Record
CVE-2017-12725- CVSS 3.0
- 5.6 Medium
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L - CVSS 2.0
- 6.8 Medium
AV:N/AC:M/Au:N/C:P/I:P/A:P - Weakness
- CWE-798
- Updated
- 2026-11-02
CVE List No score
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
- Record
CVE-2017-12725- Updated
- 2026-11-02
EUVDEUVD-2017-4264 No score
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
- Record
EUVD-2017-4264- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-chgj-p4c2-8wqj
5.6
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless...
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
- Record
GHSA-chgj-p4c2-8wqj- CVSS 3.0
- 5.6 Medium
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L - Weakness
- CWE-798
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2017-12725 is a medium vulnerability (CVSS 5.6) in Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump, published February 15, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.83%, the 57th percentile. No source names a fixed version.