Every CVE Search/ Get API key

CVE-2018-14781

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

5.3 Medium

Medtronic MiniMed MMT

devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
High
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
None
Integrity
High
Availability
None

Exploit likelihood

0.71%Up 0.1 points since Apr 14, 2021
90%50%10%1%

What each source reports

CVE List Score used 5.3

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Medtronic MiniMed MMT

devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Record
CVE-2018-14781
CVSS 3.1
5.3 MediumCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Weakness
CWE-294
Updated
2026-11-02
EUVDEUVD-2018-6665 5.3

Medtronic MiniMed MMT

devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Record
EUVD-2018-6665
CVSS 3.1
5.3 MediumCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Updated
2026-11-02
GitHub AdvisoriesGHSA-cv4q-4xgp-cr7m 5.3

Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723...

Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Record
GHSA-cv4q-4xgp-cr7m
CVSS 3.0
5.3 MediumCVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Weakness
CWE-287, CWE-294
Updated
2026-11-02
NVD 5.3

Medtronic MiniMed MMT

devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Record
CVE-2018-14781
CVSS 3.1
5.3 MediumCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
CVSS 3.0
5.3 MediumCVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
CVSS 2.0
2.9 LowAV:A/AC:M/Au:N/C:N/I:P/A:N
Weakness
CWE-294, CWE-287
Updated
2026-11-02

Affected products

Show 7 moreShow less

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2018-14781 is a medium vulnerability (CVSS 5.3) in Medtronic Mmt – 511 Pump Paradigm, published August 13, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.71%, the 53rd percentile. No source names a fixed version.

References