CVE-2018-4846
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems /...
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
NVD Score used
9.8
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
- Record
CVE-2018-4846- CVSS 3.0
- 9.8 Critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - CVSS 2.0
- 10.0 High
AV:N/AC:L/Au:N/C:C/I:C/A:C - Weakness
- CWE-798
- Updated
- 2026-11-02
CVE List No score
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
- Record
CVE-2018-4846- Weakness
- CWE-798
- Updated
- 2026-11-02
EUVDEUVD-2018-16632 No score
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
- Record
EUVD-2018-16632- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-vcx8-x3m9-gr6g
9.8
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems /...
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
- Record
GHSA-vcx8-x3m9-gr6g- CVSS 3.0
- 9.8 Critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-798
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short