CVE-2018-7510
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software...
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
NVD Score used
9.8
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
- Record
CVE-2018-7510- CVSS 3.0
- 9.8 Critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - CVSS 2.0
- 5.0 Medium
AV:N/AC:L/Au:N/C:P/I:N/A:N - Weakness
- CWE-256, CWE-522
- Updated
- 2026-11-02
CVE List No score
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
- Record
CVE-2018-7510- Weakness
- CWE-256
- Updated
- 2026-11-02
EUVDEUVD-2018-19239 No score
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
- Record
EUVD-2018-19239- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-8rp6-8jgg-qp66
9.8
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software...
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.
- Record
GHSA-8rp6-8jgg-qp66- CVSS 3.0
- 9.8 Critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-522
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2018-7510 is a critical vulnerability (CVSS 9.8) in Beaconmedaes Scroll Medical Air Systems Firmware, published June 6, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.41%, the 72nd percentile. No source names a fixed version.