BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that...

4.6 Medium

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Attack profile

How easy it is to exploit

Attack vector
Physical
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

0.36%Down 0.1 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 4.6

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Record
CVE-2019-18256
CVSS 3.1
4.6 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0
2.1 LowAV:L/AC:L/Au:N/C:P/I:N/A:N
Weakness
CWE-257, CWE-522
Updated
2026-11-02
CVE List No score

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Record
CVE-2019-18256
Weakness
CWE-257
Updated
2026-11-02
EUVDEUVD-2019-8054 No score

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Record
EUVD-2019-8054
Updated
2026-11-03
GitHub AdvisoriesGHSA-qhvv-89m3-h89m No score

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that...

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Record
GHSA-qhvv-89m3-h89m
Weakness
CWE-522
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2019-18256 is a medium vulnerability (CVSS 4.6) in Biotronik Cardiomessenger Ii-s Gsm Firmware, published June 29, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.36%, the 29th percentile. No source names a fixed version.