Every CVE Search/ Get API key

CVE-2019-25720

Dräger SC Monitoring Devices DoS via Malformed Network Packet

7.1 High

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
None
Integrity
None
Availability
High

Exploit likelihood

0.20%Up 0.2 points since Jun 4, 2026
90%50%10%1%

What each source reports

CVE List Score used 7.1

Dräger SC Monitoring Devices DoS via Malformed Network Packet

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Record
CVE-2019-25720
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness
CWE-1286
Updated
2026-11-02
EUVDEUVD-2019-20161 7.1

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Record
EUVD-2019-20161
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Updated
2026-11-02
GitHub AdvisoriesGHSA-4wcc-pjv3-vvw7 7.1

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a...

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Record
GHSA-4wcc-pjv3-vvw7
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness
CWE-1286
Updated
2026-11-02
NVD 7.1

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Record
CVE-2019-25720
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness
CWE-1286
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2019-25720 is a high vulnerability (CVSS 7.1) in Dräger Sc 6002xl, published June 3, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.20%, the 9th percentile. No source names a fixed version.

References