Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface

6.3 Medium

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
High
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
None
Integrity
None
Availability
Low

Exploit likelihood

0.24%Up 0.2 points since Jun 3, 2026
90%50%10%1%

What each source reports

CVE List Score used 6.3

Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Record
CVE-2019-25723
CVSS 4.0
6.3 MediumCVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
CVSS 3.1
4.0 MediumCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Weakness
CWE-1286
Updated
2026-11-02
EUVDEUVD-2019-20159 6.3

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Record
EUVD-2019-20159
CVSS 4.0
6.3 MediumCVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
Updated
2026-11-02
GitHub AdvisoriesGHSA-9p7r-jmp9-7hw2 6.3

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling...

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Record
GHSA-9p7r-jmp9-7hw2
CVSS 4.0
6.3 MediumCVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
4.0 MediumCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Weakness
CWE-1286
Updated
2026-11-02
NVD 6.3

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Record
CVE-2019-25723
CVSS 4.0
6.3 MediumCVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
4.0 MediumCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Weakness
CWE-1286
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2019-25723 is a medium vulnerability (CVSS 6.3) in Dräger Perseus A500, published June 2, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.24%, the 14th percentile. No source names a fixed version.

References

MoveEnter Open