Every CVE Search/ Get API key

CVE-2019-5024

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule...

7.6 High

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

Attack profile

How easy it is to exploit

Attack vector
Physical
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.47%Up 0.1 points since Apr 14, 2021
90%50%10%1%

What each source reports

CVE List Score used 7.6

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

Record
CVE-2019-5024
CVSS 3.0
7.6 HighCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-693
Updated
2026-11-02
EUVDEUVD-2019-14631 7.6

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

Record
EUVD-2019-14631
CVSS 3.0
7.6 HighCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Updated
2026-11-03
GitHub AdvisoriesGHSA-fcp8-jrh7-67m9 7.6

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule...

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

Record
GHSA-fcp8-jrh7-67m9
CVSS 3.1
7.6 HighCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Updated
2026-11-02
NVD 7.6

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

Record
CVE-2019-5024
CVSS 3.1
7.6 HighCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CVSS 3.0
7.6 HighCVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CVSS 2.0
7.2 HighAV:L/AC:L/Au:N/C:C/I:C/A:C
Weakness
CWE-693
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2019-5024 is a high vulnerability (CVSS 7.6) in Capsule Technologies Smartlinx Neuron 2, published April 11, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.47%, the 41st percentile. No source names a fixed version.

References

MoveEnter Open