CVE-2020-12036
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not...
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- None
- Availability
- None
Exploit likelihood
What each source reports
NVD Score used
7.5
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
- Record
CVE-2020-12036- CVSS 3.1
- 7.5 High
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N - CVSS 2.0
- 5.0 Medium
AV:N/AC:L/Au:N/C:P/I:N/A:N - Weakness
- CWE-319
- Updated
- 2026-11-02
CVE List No score
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
- Record
CVE-2020-12036- Weakness
- CWE-319
- Updated
- 2026-11-02
EUVDEUVD-2020-4352 No score
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
- Record
EUVD-2020-4352- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-cx3p-ghw7-pr4c No score
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not...
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
- Record
GHSA-cx3p-ghw7-pr4c- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2020-12036 is a high vulnerability (CVSS 7.5) in Baxter Prismaflex Firmware, published June 29, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.50%, the 42nd percentile. No source names a fixed version.