Every CVE Search/ Get API key

CVE-2020-12036

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not...

7.5 High

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

0.50%Up 0.1 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 7.5

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Record
CVE-2020-12036
CVSS 3.1
7.5 HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0
5.0 MediumAV:N/AC:L/Au:N/C:P/I:N/A:N
Weakness
CWE-319
Updated
2026-11-02
CVE List No score

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Record
CVE-2020-12036
Weakness
CWE-319
Updated
2026-11-02
EUVDEUVD-2020-4352 No score

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Record
EUVD-2020-4352
Updated
2026-11-03
GitHub AdvisoriesGHSA-cx3p-ghw7-pr4c No score

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not...

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Record
GHSA-cx3p-ghw7-pr4c
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2020-12036 is a high vulnerability (CVSS 7.5) in Baxter Prismaflex Firmware, published June 29, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.50%, the 42nd percentile. No source names a fixed version.

References