Every CVE Search/ Get API key

CVE-2020-12040

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System...

9.8 Critical

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.95%Up 0.5 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 9.8

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Record
CVE-2020-12040
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
5.0 MediumAV:N/AC:L/Au:N/C:P/I:N/A:N
Weakness
CWE-319
Updated
2026-11-02
CVE List No score

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Record
CVE-2020-12040
Weakness
CWE-319
Updated
2026-11-02
EUVDEUVD-2020-4356 No score

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Record
EUVD-2020-4356
Updated
2026-11-03
GitHub AdvisoriesGHSA-mr8w-rm42-6ph4 No score

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System...

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Record
GHSA-mr8w-rm42-6ph4
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2020-12040 is a critical vulnerability (CVSS 9.8) in Baxter Sigma Spectrum Infusion System Firmware, published June 29, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.95%, the 61st percentile. No source names a fixed version.

References

MoveEnter Open