The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface,...

9.4 Critical

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
Low
Availability
High

Exploit likelihood

1.41%Up 1.0 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 9.4

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Record
CVE-2020-12041
CVSS 3.1
9.4 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:H
CVSS 2.0
7.5 HighAV:N/AC:L/Au:N/C:P/I:P/A:P
Weakness
CWE-732
Updated
2026-11-02
CVE List No score

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Record
CVE-2020-12041
Weakness
CWE-732
Updated
2026-11-02
EUVDEUVD-2020-4357 No score

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Record
EUVD-2020-4357
Updated
2026-11-03
GitHub AdvisoriesGHSA-gvrr-pwvw-5j3q No score

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface,...

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Record
GHSA-gvrr-pwvw-5j3q
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2020-12041 is a critical vulnerability (CVSS 9.4) in Baxter Sigma Spectrum Infusion System Firmware, published June 29, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.41%, the 72nd percentile. No source names a fixed version.