The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a...

9.8 Critical

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

1.68%Up 1.3 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 9.8

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.

Record
CVE-2020-12045
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
7.5 HighAV:N/AC:L/Au:N/C:P/I:P/A:P
Weakness
CWE-259, CWE-798
Updated
2026-11-02
CVE List No score

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.

Record
CVE-2020-12045
Weakness
CWE-259
Updated
2026-11-02
EUVDEUVD-2020-4361 No score

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.

Record
EUVD-2020-4361
Updated
2026-11-03
GitHub AdvisoriesGHSA-gxx4-h2mv-92pp No score

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a...

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.

Record
GHSA-gxx4-h2mv-92pp
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2020-12045 is a critical vulnerability (CVSS 9.8) in Baxter Sigma Spectrum Infusion System Firmware, published June 29, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.68%, the 77th percentile. No source names a fixed version.