Every CVE Search/ Get API key

CVE-2020-15486

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is...

6.5 Medium

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

0.51%Up 0.1 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 6.5

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Record
CVE-2020-15486
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0
3.3 LowAV:A/AC:L/Au:N/C:P/I:N/A:N
Updated
2026-11-02
CVE List No score

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Record
CVE-2020-15486
Updated
2026-11-02
EUVDEUVD-2020-7479 No score

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Record
EUVD-2020-7479
Updated
2026-11-03
GitHub AdvisoriesGHSA-xvmj-frj8-3xq7 No score

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is...

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Record
GHSA-xvmj-frj8-3xq7
Weakness
CWE-200
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2020-15486 is a medium vulnerability (CVSS 6.5) in Drtrust Electrocardiogram Pen Firmware, published August 26, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.51%, the 43rd percentile. No source names a fixed version.

References

MoveEnter Open