Every CVE Search/ Get API key

CVE-2020-27276

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the...

5.7 Medium

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
Required

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

0.51%Up 0.1 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 5.7

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.

Record
CVE-2020-27276
CVSS 3.1
5.7 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N
CVSS 2.0
2.9 LowAV:A/AC:M/Au:N/C:P/I:N/A:N
Weakness
CWE-290
Updated
2026-11-02
CVE List No score

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.

Record
CVE-2020-27276
Weakness
CWE-290
Updated
2026-11-02
EUVDEUVD-2020-19789 No score

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.

Record
EUVD-2020-19789
Updated
2026-11-03
GitHub AdvisoriesGHSA-hpq4-rvv8-g82j No score

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the...

SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.

Record
GHSA-hpq4-rvv8-g82j
Weakness
CWE-290
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2020-27276 is a medium vulnerability (CVSS 5.7) in Sooil Anydana-a Firmware, published January 19, 2021. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.51%, the 43rd percentile. No source names a fixed version.

References

MoveEnter Open