CVE-2020-27276
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the...
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
Attack profile
How easy it is to exploit
- Attack vector
- Adjacent network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- Required
What an attacker can affect
- Confidentiality
- High
- Integrity
- None
- Availability
- None
Exploit likelihood
What each source reports
NVD Score used
5.7
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
- Record
CVE-2020-27276- CVSS 3.1
- 5.7 Medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N - CVSS 2.0
- 2.9 Low
AV:A/AC:M/Au:N/C:P/I:N/A:N - Weakness
- CWE-290
- Updated
- 2026-11-02
CVE List No score
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
- Record
CVE-2020-27276- Weakness
- CWE-290
- Updated
- 2026-11-02
EUVDEUVD-2020-19789 No score
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
- Record
EUVD-2020-19789- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-hpq4-rvv8-g82j No score
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the...
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
- Record
GHSA-hpq4-rvv8-g82j- Weakness
- CWE-290
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short