In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior,...
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
NVD Score used
10.0
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
- Record
CVE-2020-6963- CVSS 3.1
- 10.0 Critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H - CVSS 2.0
- 10.0 High
AV:N/AC:L/Au:N/C:C/I:C/A:C - Weakness
- CWE-798, CWE-20
- Updated
- 2026-11-02
CVE List No score
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
- Record
CVE-2020-6963- Weakness
- CWE-798
- Updated
- 2026-11-02
EUVDEUVD-2020-28103 No score
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
- Record
EUVD-2020-28103- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-xrpc-f6j4-wmx4 No score
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior,...
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
- Record
GHSA-xrpc-f6j4-wmx4- Updated
- 2026-11-02
Affected products
- gehealthcare apexpro telemetry server firmware
- gehealthcare carescape central station mai700 firmware
- gehealthcare carescape central station mas700 firmware
- gehealthcare carescape telemetry server mp100r firmware
- gehealthcare clinical information center mp100d firmware
- gehealthcare clinical information center mp100r firmware
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2020-6963 is a critical vulnerability (CVSS 10.0) in Gehealthcare Apexpro Telemetry Server Firmware, published January 24, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 2.75%, the 86th percentile. No source names a fixed version.