CVE-2020-7590
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by...
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
Attack profile
How easy it is to exploit
- Attack vector
- Physical
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
NVD Score used
6.8
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
- Record
CVE-2020-7590- CVSS 3.1
- 6.8 Medium
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - CVSS 2.0
- 4.6 Medium
AV:L/AC:L/Au:N/C:P/I:P/A:P - Weakness
- CWE-259
- Updated
- 2026-11-02
CVE List No score
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
- Record
CVE-2020-7590- Weakness
- CWE-259
- Updated
- 2026-11-02
EUVDEUVD-2020-28715 No score
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
- Record
EUVD-2020-28715- Updated
- 2026-11-03
GitHub AdvisoriesGHSA-cqp5-5522-8hhp No score
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by...
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
- Record
GHSA-cqp5-5522-8hhp- Weakness
- CWE-259
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| facts | ||
| aliases | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2020-7590 is a medium vulnerability (CVSS 6.8) in Siemens Dca Vantage Analyzer, published October 13, 2020. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.35%, the 29th percentile. No source names a fixed version.