Every CVE Search/ Get API key

CVE-2021-37160

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus...

9.8 Critical

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

8.23%Up 7.2 points since Aug 3, 2021
90%50%10%1%

What each source reports

NVD Score used 9.8

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Record
CVE-2021-37160
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
7.5 HighAV:N/AC:L/Au:N/C:P/I:P/A:P
Weakness
CWE-347
Updated
2026-11-02
CVE List No score

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Record
CVE-2021-37160
Updated
2026-11-02
EUVDEUVD-2021-23734 No score

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Record
EUVD-2021-23734
Updated
2026-11-03
GitHub AdvisoriesGHSA-3vvw-g7qx-ffjj 9.8

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus...

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Record
GHSA-3vvw-g7qx-ffjj
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-347
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2021-37160 is a critical vulnerability (CVSS 9.8) in Swisslog-healthcare Hmi-3 Control Panel Firmware, published August 2, 2021. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 8.23%, the 94th percentile. No source names a fixed version.

References

MoveEnter Open