Every CVE Search/ Get API key

CVE-2021-38396

Missing Support Integrity Check for Boston Scientific Zoom Latitude

6.5 Medium

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Attack profile

How easy it is to exploit

Attack vector
Physical
Complexity
Low
Privileges needed
None
User interaction
Required

What an attacker can affect

Confidentiality
Low
Integrity
High
Availability
Low

Exploit likelihood

0.17%Down 0.2 points since Oct 5, 2021
90%50%10%1%

What each source reports

CVE List Score used 6.5

Missing Support Integrity Check for Boston Scientific Zoom Latitude

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Record
CVE-2021-38396
CVSS 3.1
6.5 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L
Weakness
CWE-353
Updated
2026-11-02
EUVDEUVD-2021-24848 6.5

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Record
EUVD-2021-24848
CVSS 3.1
6.5 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L
Updated
2026-11-03
GitHub AdvisoriesGHSA-phv8-rwj8-6jj2 6.8

The programmer installation utility does not perform a cryptographic authenticity or integrity...

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Record
GHSA-phv8-rwj8-6jj2
CVSS 3.1
6.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-345
Updated
2026-11-02
NVD 6.5

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Record
CVE-2021-38396
CVSS 3.1
6.5 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L
CVSS 3.1
6.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.0
4.6 MediumAV:L/AC:L/Au:N/C:P/I:P/A:P
Weakness
CWE-353, CWE-345
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliases, publishedEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2021-38396 is a medium vulnerability (CVSS 6.5) in Boston Scientific Zoom Latitude, published October 4, 2021. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.17%, the 6th percentile. No source names a fixed version.