Every CVE Search/ Get API key

CVE-2022-36385

Contec Health CMS8000

6.8 Medium

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Attack profile

How easy it is to exploit

Attack vector
Physical
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.41%Down 0.5 points since Sep 14, 2022
90%50%10%1%

What each source reports

CVE List Score used 6.8

Contec Health CMS8000

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Record
CVE-2022-36385
CVSS 3.1
6.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-284
Updated
2026-11-02
EUVDEUVD-2022-39098 6.8

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Record
EUVD-2022-39098
CVSS 3.1
6.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Updated
2026-11-02
GitHub AdvisoriesGHSA-gvx4-5vjq-6x82 6.8

A threat actor with momentary access to the device can plug in a USB drive and perform a...

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Record
GHSA-gvx4-5vjq-6x82
CVSS 3.1
6.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Updated
2026-11-02
NVD 6.8

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Record
CVE-2022-36385
CVSS 3.1
6.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-284
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2022-36385 is a medium vulnerability (CVSS 6.8) in Contec Health Cms8000 Contec Icu Ccu Vital Signs Patient Monitor, published September 13, 2022. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.41%, the 35th percentile. No source names a fixed version.

References

MoveEnter Open