Every CVE Search/ Get API key

CVE-2022-4992

Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering

8.8 High

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
Low
Integrity
Low
Availability
High

Exploit likelihood

0.16%Up 0.1 points since Jun 3, 2026
90%50%10%1%

What each source reports

CVE List Score used 8.8

Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Record
CVE-2022-4992
CVSS 4.0
8.8 HighCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
CVSS 3.1
8.6 HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Weakness
CWE-345
Updated
2026-11-02
EUVDEUVD-2022-55996 8.8

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Record
EUVD-2022-55996
CVSS 4.0
8.8 HighCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
Updated
2026-11-02
GitHub AdvisoriesGHSA-6jx8-hc3v-28hr 8.8

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1,...

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Record
GHSA-6jx8-hc3v-28hr
CVSS 4.0
8.8 HighCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
8.6 HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Weakness
CWE-345
Updated
2026-11-02
NVD 8.8

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Record
CVE-2022-4992
CVSS 4.0
8.8 HighCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
8.6 HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Weakness
CWE-345
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2022-4992 is a high vulnerability (CVSS 8.8) in Dräger Infinity Acute Care System, published June 2, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.16%, the 5th percentile. No source names a fixed version.

References

MoveEnter Open