Every CVE Search/ Get API key

CVE-2024-48966

Life2000 service tools for test and calibration do not support user authentication

10.0 Critical

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.69%Up 0.7 points since Nov 15, 2024
90%50%10%1%

What each source reports

CVE List Score used 10.0

Life2000 service tools for test and calibration do not support user authentication

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Record
CVE-2024-48966
CVSS 3.1
10.0 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-306
Updated
2026-11-02
EUVDEUVD-2024-43153 10.0

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Record
EUVD-2024-43153
CVSS 3.1
10.0 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Updated
2026-11-03
GitHub AdvisoriesGHSA-285j-vfp9-5w9w 10.0

The software tools used by service personnel to test & calibrate the ventilator do not support...

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Record
GHSA-285j-vfp9-5w9w
CVSS 3.1
10.0 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-306
Updated
2026-11-02
NVD 10.0

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Record
CVE-2024-48966
CVSS 3.1
10.0 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-306
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliases, publishedEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2024-48966 is a critical vulnerability (CVSS 10.0) in Baxter Life2000 Ventilation System, published November 14, 2024. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.69%, the 53rd percentile. No source names a fixed version.

References

MoveEnter Open