CVE-2024-48974
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Attack profile
How easy it is to exploit
- Attack vector
- Local
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
CVE List Score used
9.3
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
- Record
CVE-2024-48974- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H - Weakness
- CWE-494
- Updated
- 2026-11-02
EUVDEUVD-2024-43158
9.3
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
- Record
EUVD-2024-43158- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H - Updated
- 2026-11-03
GitHub AdvisoriesGHSA-rfvx-872w-6rpr
9.3
The ventilator does not perform proper file integrity checks when adopting firmware updates. This...
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
- Record
GHSA-rfvx-872w-6rpr- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H - Weakness
- CWE-494
- Updated
- 2026-11-02
NVD
9.3
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
- Record
CVE-2024-48974- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H - Weakness
- CWE-494
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| aliases, published | EUVD | |
| first indexed | CVE List, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2024-48974 is a critical vulnerability (CVSS 9.3) in Baxter Life2000 Ventilation System, published November 14, 2024. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.15%, the 4th percentile. No source names a fixed version.