CVE-2025-0626
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- Passive
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
CVE List Score used
7.7
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
- Record
CVE-2025-0626- CVSS 4.0
- 7.7 High
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N - CVSS 3.1
- 7.5 High
CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H - Weakness
- CWE-912
- Updated
- 2026-11-02
EUVDEUVD-2025-1797
7.7
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
- Record
EUVD-2025-1797- CVSS 4.0
- 7.7 High
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N - Updated
- 2026-11-02
GitHub AdvisoriesGHSA-p2j3-fm26-85wv
7.7
The affected product sends out remote access requests to a hard-coded IP address, bypassing...
The affected product sends out remote access requests to a hard-coded IP address, bypassing existing device network settings to do so. This could serve as a backdoor and lead to a malicious actor being able to upload and overwrite files on the device.
- Record
GHSA-p2j3-fm26-85wv- CVSS 4.0
- 7.7 High
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X - CVSS 3.1
- 7.5 High
CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H - Weakness
- CWE-912
- Updated
- 2026-11-02
NVD
7.7
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
- Record
CVE-2025-0626- CVSS 4.0
- 7.7 High
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X - CVSS 3.1
- 7.5 High
CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H - Weakness
- CWE-912
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| first indexed | CVE List, EUVD, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2025-0626 is a high vulnerability (CVSS 7.7) in Contec Health Cms8000 Patient Monitor, published January 30, 2025. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.14%, the 66th percentile. No source names a fixed version.
References
cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china
nvd.nist.gov/vuln/detail/CVE-2025-0626
github.com/advisories/GHSA-p2j3-fm26-85wv