CVE-2026-0611
Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- High
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
CVE List Score used
9.2
Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
- Record
CVE-2026-0611- CVSS 4.0
- 9.2 Critical
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N - CVSS 3.1
- 9.8 Critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-306
- Updated
- 2026-11-02
EUVDEUVD-2026-33974
9.2
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
- Record
EUVD-2026-33974- CVSS 4.0
- 9.2 Critical
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N - Updated
- 2026-11-02
GitHub AdvisoriesGHSA-344f-p6q5-rw6q
9.2
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an...
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
- Record
GHSA-344f-p6q5-rw6q- CVSS 4.0
- 9.2 Critical
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X - CVSS 3.1
- 9.8 Critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-306
- Updated
- 2026-11-02
NVD
9.2
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
- Record
CVE-2026-0611- CVSS 4.0
- 9.2 Critical
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X - CVSS 3.1
- 9.8 Critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-306
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| first indexed | CVE List, EUVD, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2026-0611 is a critical vulnerability (CVSS 9.2) in Netiq Corporation Sentinel, published June 2, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.66%, the 52nd percentile. No source names a fixed version.
References
spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf
spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel
vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting
nvd.nist.gov/vuln/detail/CVE-2026-0611
github.com/advisories/GHSA-344f-p6q5-rw6q