Every CVE Search/ Get API key

CVE-2026-18844

Pulsetto Vagus Nerve Stimulator Hidden Functionality

7.2 High

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
None
Integrity
High
Availability
High

Exploit likelihood

0.20%Up 0.1 points since Aug 12, 2026
90%50%10%1%

What each source reports

CVE List Score used 7.2

Pulsetto Vagus Nerve Stimulator Hidden Functionality

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.

Record
CVE-2026-18844
CVSS 4.0
7.2 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N
CVSS 3.1
8.1 HighCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
Weakness
CWE-912
Updated
2026-11-02
EUVDEUVD-2026-56936 7.2

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.

Record
EUVD-2026-56936
CVSS 4.0
7.2 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N
Updated
2026-11-02
GitHub AdvisoriesGHSA-qhgm-pqrh-9j38 7.2

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its...

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.

Record
GHSA-qhgm-pqrh-9j38
CVSS 4.0
7.2 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
8.1 HighCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
Weakness
CWE-912
Updated
2026-11-02
NVD 7.2

The firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.

Record
CVE-2026-18844
CVSS 4.0
7.2 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
8.1 HighCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
Weakness
CWE-912
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2026-18844 is a high vulnerability (CVSS 7.2) in Pulsetto Vagus Nerve Stimulator, published August 11, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.20%, the 10th percentile. No source names a fixed version.

References

MoveEnter Open