Every CVE Search/ Get API key

CVE-2026-66098

Mira Hormone Monitor, Mira Android App Missing authentication for critical function

7.1 High

The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
None
Integrity
None
Availability
High

Exploit likelihood

0.32%Up 0.1 points since Aug 12, 2026
90%50%10%1%

What each source reports

CVE List Score used 7.1

Mira Hormone Monitor, Mira Android App Missing authentication for critical function

The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.

Record
CVE-2026-66098
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness
CWE-306
Updated
2026-11-02
EUVDEUVD-2026-57007 7.1

The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.

Record
EUVD-2026-57007
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Updated
2026-11-02
GitHub AdvisoriesGHSA-5pc2-vj8c-ggcr 7.1

The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without...

The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.

Record
GHSA-5pc2-vj8c-ggcr
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness
CWE-306
Updated
2026-11-02
NVD 7.1

The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.

Record
CVE-2026-66098
CVSS 4.0
7.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness
CWE-306
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2026-66098 is a high vulnerability (CVSS 7.1) in Quanovate Tech Inc. (operating As Mira / Mira Care) Mira Android App, published August 11, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.32%, the 25th percentile. No source names a fixed version.

References

MoveEnter Open