Every CVE Search/ Get API key

CVE-2026-7251

Eppendorf BioFlo 320 Use of hard-coded password

9.3 Critical

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Attack profile

How easy it is to exploit

Attack vector
Network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.50%Up 0.4 points since May 27, 2026
90%50%10%1%

What each source reports

CVE List Score used 9.3

Eppendorf BioFlo 320 Use of hard-coded password

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Record
CVE-2026-7251
CVSS 4.0
9.3 CriticalCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-259
Updated
2026-11-02
EUVDEUVD-2026-31912 9.3

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Record
EUVD-2026-31912
CVSS 4.0
9.3 CriticalCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Updated
2026-11-02
GitHub AdvisoriesGHSA-v6xg-wr2p-xrj3 9.3

Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote...

Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Record
GHSA-v6xg-wr2p-xrj3
CVSS 4.0
9.3 CriticalCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-259
Updated
2026-11-02
NVD 9.3

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Record
CVE-2026-7251
CVSS 4.0
9.3 CriticalCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
CVSS 3.1
9.8 CriticalCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness
CWE-259
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2026-7251 is a critical vulnerability (CVSS 9.3) in Eppendorf Bioflo 320, published May 26, 2026. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.50%, the 42nd percentile. No source names a fixed version.

References

MoveEnter Open