Every CVESearch/Get API key

CVE-2018-18563

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number...

9.6 Critical

CVE-2018-18563 is a critical vulnerability (CVSS 9.6) in Roche Accu-chek Inform Ii Firmware, published November 20, 2018. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.97%, the 62nd percentile. No source names a fixed version.

What does this mean?
Severity
9.6 CriticalCVSS 3.0, NVD
Exploit chance
0.97%EPSS, 62nd percentile
Exploitation
None knownNot on CISA’s or ENISA’s lists
Fixed in
No fix listedNo source names one

Description

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
High
Availability
High

Exploit likelihood

0.97%Down 0.1 points since Apr 14, 2021
90%50%10%1%

Affected products

What each source reports

NVD Score used9.6

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Record
CVE-2018-18563
CVSS 3.0
9.6 CriticalCVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CVSS 2.0
8.3 HighAV:A/AC:L/Au:N/C:C/I:C/A:C
Weakness
CWE-434
Updated
2026-11-02
CVE ListNo score

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Record
CVE-2018-18563
Updated
2026-11-02
EUVDEUVD-2018-10283No score

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Record
EUVD-2018-10283
Updated
2026-11-03
GitHub AdvisoriesGHSA-fvc7-cwmq-hcv29.6

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number...

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Record
GHSA-fvc7-cwmq-hcv2
CVSS 3.0
9.6 CriticalCVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Weakness
CWE-434
Updated
2026-11-02

History

History
RecordedChangedSources
facts
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

References

MoveEnter Open