CVE-2014-5406
Hospira LifeCare PCA Infusion System
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Exploit likelihood
What each source reports
CVE List Score used
7.6
Hospira LifeCare PCA Infusion System
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
- Record
CVE-2014-5406- CVSS 2.0
- 7.6
AV:N/AC:H/Au:N/C:C/I:C/A:C - Weakness
- CWE-345
- Updated
- 2026-11-02
EUVDEUVD-2014-5294
7.6
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
- Record
EUVD-2014-5294- CVSS 2.0
- 7.6 High
AV:N/AC:H/Au:N/C:C/I:C/A:C - Updated
- 2026-11-02
GitHub AdvisoriesGHSA-vxf4-qh89-w2r6 No score
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated...
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
- Record
GHSA-vxf4-qh89-w2r6- Weakness
- CWE-345
- Updated
- 2026-11-02
NVD
7.6
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
- Record
CVE-2014-5406- CVSS 2.0
- 7.6 High
AV:N/AC:H/Au:N/C:C/I:C/A:C - CVSS 2.0
- 9.3 High
AV:N/AC:M/Au:N/C:C/I:C/A:C - Weakness
- CWE-345
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| first indexed | CVE List, EUVD, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2014-5406 is a vulnerability (CVSS 7.6) in Hospira Lifecare Pca Infusion System, published July 6, 2015. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 1.22%, the 68th percentile. No source names a fixed version.
References
cisa.gov/news-events/ics-advisories/icsa-15-125-01
fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities
github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
ics-cert.us-cert.gov/advisories/ICSA-15-125-01
nvd.nist.gov/vuln/detail/CVE-2014-5406
github.com/advisories/GHSA-vxf4-qh89-w2r6