Every CVE Search/ Get API key

CVE-2019-6538

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

9.3 Critical

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
None
Integrity
High
Availability
High

Exploit likelihood

0.90%Up 0.3 points since Apr 14, 2021
90%50%10%1%

What each source reports

CVE List Score used 9.3

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Record
CVE-2019-6538
CVSS 3.1
9.3 CriticalCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
Weakness
CWE-284
Updated
2026-11-02
EUVDEUVD-2019-16097 9.3

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Record
EUVD-2019-16097
CVSS 3.1
9.3 CriticalCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
Updated
2026-11-02
GitHub AdvisoriesGHSA-56wm-53rq-h4x6 6.5

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and...

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Record
GHSA-56wm-53rq-h4x6
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
Weakness
CWE-284, CWE-306
Updated
2026-11-02
NVD 9.3

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Record
CVE-2019-6538
CVSS 3.1
9.3 CriticalCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
CVSS 3.1
6.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
CVSS 2.0
3.3 LowAV:A/AC:L/Au:N/C:N/I:P/A:N
Weakness
CWE-284, CWE-306, CWE-862
Updated
2026-11-02

Affected products

Show 32 moreShow less

History

History
RecordedChangedSources
first indexedCVE List, EUVD, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2019-6538 is a critical vulnerability (CVSS 9.3) in Medtronic Amplia Crt-d, published March 25, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.90%, the 59th percentile. No source names a fixed version.

References

MoveEnter Open