CVE-2019-6538
Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Attack profile
How easy it is to exploit
- Attack vector
- Adjacent network
- Complexity
- Low
- Privileges needed
- None
- User interaction
- None
What an attacker can affect
- Confidentiality
- None
- Integrity
- High
- Availability
- High
Exploit likelihood
What each source reports
CVE List Score used
9.3
Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
- Record
CVE-2019-6538- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H - Weakness
- CWE-284
- Updated
- 2026-11-02
EUVDEUVD-2019-16097
9.3
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
- Record
EUVD-2019-16097- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H - Updated
- 2026-11-02
GitHub AdvisoriesGHSA-56wm-53rq-h4x6
6.5
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and...
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
- Record
GHSA-56wm-53rq-h4x6- CVSS 3.1
- 6.5 Medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N - Weakness
- CWE-284, CWE-306
- Updated
- 2026-11-02
NVD
9.3
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
- Record
CVE-2019-6538- CVSS 3.1
- 9.3 Critical
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H - CVSS 3.1
- 6.5 Medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N - CVSS 2.0
- 3.3 Low
AV:A/AC:L/Au:N/C:N/I:P/A:N - Weakness
- CWE-284, CWE-306, CWE-862
- Updated
- 2026-11-02
Affected products
- medtronic amplia crt-d
- medtronic carelink 2090 programmer
- medtronic carelink monitor
- medtronic claria crt-d
- medtronic compia crt-d
- medtronic concerto crt-d
- medtronic concerto ii crt-d
- medtronic conexus radio frequency telemetry protocol
- medtronic consulta crt-d
- medtronic evera icd
- medtronic maximo ii crt-d
- medtronic mirro icd
Show 32 moreShow less
- medtronic mycarelink monitor
- medtronic nayamed nd icd
- medtronic primo icd
- medtronic secura icd
- medtronic virtuoso icd
- medtronic virtuoso ii icd
- medtronic visia af icd
- medtronic viva crt-d
- medtronic amplia crt-d firmware
- medtronic brava crt-d
- medtronic carelink 2090 firmware
- medtronic carelink monitor firmware
- medtronic claria crt-d firmware
- medtronic compia crt-d firmware
- medtronic concerto crt-d firmware
- medtronic concerto ii crt-d firmware
- medtronic consulta crt-d firmware
- medtronic evera icd firmware
- medtronic maximo ii icd
- medtronic maximo ii crt-d and lcd firmware
- medtronic mirro mri icd
- medtronic mirro icd firmware
- medtronic mycarelink monitor firmware
- medtronic nayamed nd icd firmware
- medtronic primo icd firmware
- medtronic protecta icd, protecta crt-d
- medtronic protecta icd and crt-d firmware
- medtronic secura icd firmware
- medtronic virtuoso icd firmware
- medtronic virtuoso ii icd firmware
- medtronic visia af icd firmware
- medtronic viva crt-d firmware
History
| Recorded | Changed | Sources |
|---|---|---|
| first indexed | CVE List, EUVD, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2019-6538 is a critical vulnerability (CVSS 9.3) in Medtronic Amplia Crt-d, published March 25, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.90%, the 59th percentile. No source names a fixed version.