Every CVE Search/ Get API key

CVE-2018-19014

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version,...

6.5 Medium

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

Attack profile

How easy it is to exploit

Attack vector
Adjacent network
Complexity
Low
Privileges needed
None
User interaction
None

What an attacker can affect

Confidentiality
High
Integrity
None
Availability
None

Exploit likelihood

0.78%Up 0.2 points since Apr 14, 2021
90%50%10%1%

What each source reports

NVD Score used 6.5

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

Record
CVE-2018-19014
CVSS 3.0
6.5 MediumCVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.0
3.3 LowAV:A/AC:L/Au:N/C:P/I:N/A:N
Weakness
CWE-532
Updated
2026-11-02
CVE List No score

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

Record
CVE-2018-19014
Weakness
CWE-532
Updated
2026-11-02
EUVDEUVD-2018-10732 No score

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

Record
EUVD-2018-10732
Updated
2026-11-03
GitHub AdvisoriesGHSA-j27r-78g6-gwxh 6.5

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version,...

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

Record
GHSA-j27r-78g6-gwxh
CVSS 3.0
6.5 MediumCVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Weakness
CWE-532
Updated
2026-11-02

Affected products

History

History
RecordedChangedSources
aliasesEUVD
first indexedCVE List, GitHub Advisories, NVD

Dates are when Every CVE recorded each change. Upstream dates may differ.

In short

CVE-2018-19014 is a medium vulnerability (CVSS 6.5) in Draeger Delta Xl Firmware, published January 28, 2019. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.78%, the 56th percentile. No source names a fixed version.

References

MoveEnter Open