CVE-2023-0888
Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Attack profile
How easy it is to exploit
- Attack vector
- Network
- Complexity
- Low
- Privileges needed
- High
- User interaction
- None
What an attacker can affect
- Confidentiality
- None
- Integrity
- High
- Availability
- None
Exploit likelihood
What each source reports
CVE List Score used
4.9
Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
- Record
CVE-2023-0888- CVSS 3.1
- 4.9 Medium
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N - Weakness
- CWE-95
- Updated
- 2026-11-02
EUVDEUVD-2023-12877
4.9
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
- Record
EUVD-2023-12877- CVSS 3.1
- 4.9 Medium
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N - Updated
- 2026-11-02
GitHub AdvisoriesGHSA-9w87-qf4h-4fc6
7.2
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi...
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
- Record
GHSA-9w87-qf4h-4fc6- CVSS 3.1
- 7.2 High
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-94
- Updated
- 2026-11-02
NVD
4.9
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
- Record
CVE-2023-0888- CVSS 3.1
- 4.9 Medium
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N - CVSS 3.1
- 7.2 High
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H - Weakness
- CWE-95, CWE-94
- Updated
- 2026-11-02
Affected products
History
| Recorded | Changed | Sources |
|---|---|---|
| first indexed | CVE List, EUVD, GitHub Advisories, NVD |
Dates are when Every CVE recorded each change. Upstream dates may differ.
In short
CVE-2023-0888 is a medium vulnerability (CVSS 4.9) in B. Braun Space Battery Pack Sp With Wi-fi, published March 13, 2023. No source lists public exploit code. EPSS puts the chance of exploitation in the next 30 days at 0.60%, the 49th percentile. No source names a fixed version.